Please enter valid quantity
Please log in to add favorites.
NULL OR EMPTY CART
The new Advanced Platform design suspends the platform in the tube using three small pins in a triangular layout. This has the advantage of minimizing wall contact and reducing thermal conduction from the tube walls resulting in even and rapid heating almost exclusively from thermal radiation, an ideal platform.
The exceptionally uniform temperature distribution obtained with transverse heating significantly reduces or eliminates condensation of the sample matrix components and "memory" effects and improves the atomization efficiency for refractory elements. For improved characteristic mass with volatile and medium refractory elements such as Ag, As, Cd, Co, Cr, Cu, Pb, Se, and Tl, the THGA graphite tubes with end caps are recommended.
Features and Benefits:
NOTE: THGA graphite tubes with end caps are not recommended for refractory elements such as V or for Ba.
The tubes are compatible with PerkinElmer PinAAcle 900T, PinAAcle 900Z, AAnalyst 600, AAnalyst 800, Model 4100ZL, Model 4110ZL, Model 5100ZL, SIMAA 6000 Zeeman AA instruments.
Product Brand Name | PerkinElmer |
---|---|
Product Group | THGA Graphite Tubes |
Style | Advanced Platform |
Type | End-Capped |
Unit Size | 10 |
PerkinElmer tubes are manufactured to the highest quality specifications. Our exclusive design ensures even heating over the entire length of the tube. Performance must be reproducible from day to day with respect to sensitivity; accuracy and precision. W
The production of high-quality graphite components for atomic absorption spectroscopy requires stringent quality control. To ensure high quality and consistency, a specific high-density base graphite material has been developed for use with PerkinElmer graphite furnace systems.